Monthly Maintenance
Replace Alkaline Buffer
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-22
October 2005
6
Clip the damper assembly back into the upright position.
7
CAUTION
Pinching Hazard. When lowering the ISE module, keep fingers clear.
Lower the ISE module down into cavity and reinstall the ISE module
cover.
If the fluid level is above the MAX line:
8
Remove input line #30 (1) feeding into the damper.
• Using a clean transfer pipette, remove excess reagent until damper is
one-half full. Discard the removed fluid.
• Reconnect input line #30 to the damper, wipe dry any reagent spill on the
damper top.
Figure 10.10
9
Prime the alkaline buffer three times by selecting:
• the Utils icon from the menu bar,
• <1> Prime,
• MC [F4],
• ISE CO
2
Alkaline Buffer.
Type 3 in the Number of primes to repeat field.
Select <Start Prime>. (Prime cycles are completed when Standby appears
in the system status.)
Step
Action
, continued
(2 of 3)