System Power On/Boot
Introduction
UniCel DxC Systems Instructions For Use A13914
System Status and Commands
October 2005
Page 11-19
11
System Power On/Boot
Introduction
The first software routine executed from a power-on or reset condition is the boot-up.
Boot software places the instrument into a functional state by performing system
initialization, running boot diagnostics, and loading the operating system.
Power Up Sequence
The following table describes the power-up sequence often referred to as a hard (or
cold) boot.
Full Boot
When power has been restored or a reboot has occurred, the boot-up routine will begin
automatically. The monitor will show the Instrument Boot Status screen.
• Hardware Diagnostics are performed first. OK appears upon completion, unless an
error condition is found.
• Following successful hardware diagnostics, the ICC Boot will begin. A bar graph
appears indicating how far the ICC Boot has progressed.
• The MSMC Boot follows the ICC Boot. Again, a bar graph shows progress.
• The next area booted is the Smart Module Boot; a bar graph shows progress.
• The final step includes homing of all mechanisms, removal of any racks on the
sample carousel, and priming of the system. Upon completion of the boot sequence,
the instrument status shown on the console changes to Standby.
Following Boot-up
If a temperature error occurs immediately following system boot-up, no action is
required because the system temperature has not yet equilibrated. During temperature
equilibration, which takes approximately 30 minutes, temperature errors should be
ignored.
The Glucose sensor may drift over the first 24 to 48 hours. Verify control recovery at
4 hours and recalibrate if necessary. More frequent calibration may be required during
the first 24 to 48 hours following installation.
Step
Action
1
Turn on the monitor. Wait several seconds to start the communication.
2
Turn on the printer.
3
Turn on the console. If an external Uninterruptible Power Supply (UPS) is
installed, make sure that it is turned on.
4
Turn the instrument Main circuit breaker to the ON position.