Monthly Maintenance
Replace Alkaline Buffer
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-19
10
5
Remove reagent input line #30 (1) (line draining into damper) from the top
of the in-line filter.
Figure 10.7
6
Twist off the in-line filter (P/N 669212) from the input line and replace.
Note: When installing in-line filters, be sure the arrow on the filter is
pointing in the direction of reagent flow.
7
After installation of the filter is complete, reattach reagent input line #30
(that flows into damper) to the top of the filter.
8
Select the Rgts/Cal icon from the menu bar.
9
Select <MC Rgts>.
10
Select Load [F1].
Step
Action
, continued
(2 of 3)