Summary of Precautions
Introduction
UniCel DxC Systems Instructions For Use A13914
General Information
October 2005
Page 1-7
1
Summary of Precautions
Introduction
This section summarizes the precautions that should be taken when operating the DxC
System. Individual precautions associated with a specific procedure in this manual are
included in Caution boxes within the procedures for that task. Please read this section
and the preceding
before operating the system.
AccuSense™ Glucose Sensor Precautions
Whenever the sensor is replaced, reaction cup and stir bar cleaning is recommended.
Remove the sensor prior to removing stir bar. To prevent damage to the glucose
oxygen sensor membrane tip, do not insert stir bar removal tool, applicator stick, or
any other object into the glucose reaction cup unless the sensor has been removed. Do
not touch membrane tip of the AccuSense glucose oxygen sensor.
Air Filter Cleaning Precautions
Do not place a damp filter back on the system. Residual moisture may damage the
system.
Alkaline Buffer Stability Precautions
The alkaline buffer reagent is stable for one month on the system. However, if a color
change from pink to a lighter shade of pink should occur, replace the alkaline buffer
with a fresh bottle of reagent.
Aqueous Calibrator Precautions
Repetitive refrigeration of Synchron aqueous calibrators may facilitate crystal
formation. Once removed from refrigerated storage, these calibrators should remain at
room temperature. The calibrator once opened is stable for the period claimed in the
accompanying package insert.
Bar Code Label Precautions
A misread label can cause one sample ID to be read as another. The laboratory’s
process for printing, placing, and meeting all bar code specifications is important to
achieve highly accurate readings. Always follow the bar code label specifications to
avoid misread labels.
Beckman Coulter Microtube™ Precautions
• Beckman Coulter Microtubes™ are designed for use on specific Synchron systems.
Using the appropriate Microtube is essential for proper system operation.
• The sample height in the Microtube is critical for correct sample aspiration on all
Synchron systems.
• The use of Array
®
Microtubes™ (P/N 448163 or P/N 448162) on Synchron
Systems or the use of Synchron Microtubes (P/N 756776) on Array systems may
result in short sampling, incorrect results, and/or sample probe damage.