QC Chart (Levey-Jennings)
Introduction
UniCel DxC Systems Instructions For Use A13914
Quality Control
October 2005
Page 6-17
6
QC Chart (Levey-Jennings)
Introduction
QC Chart shows the results of a control for a specified period (default is current date)
in a graphic form, showing the position of data points relative to the assigned mean
and standard deviation. The results are listed by date and time, most recent results
first. QC Chart is available from either the hard disk or the floppy disk.
Note: Color flagging is based on assigned mean and SD at time of run. If assigned
mean and/or SD are modified, flag color will not change.
Accessing QC Charts
Follow the steps below to access QC charts.
Step
Action
1
Select the QC icon from the menu bar.
2
From the QC screen, select a currently defined control or type the item
number of the desired control for viewing.
3
Select Chart [F7].
4
Type a Start and End date for the QC Chart in the date range fields or press
[Enter] to default to the current date. Press the [Tab] key to move between
data fields.
Note: If no data exists for the dates entered, the message "No Data
Available" appears.
(1 of 2)