Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-66
October 2005
0x13A00023
10
QC Database is
full. The oldest
result has been
overwritten.
WLM: QC DB is full.
The oldest result for
control: control name /
Chem: chem name has
been overwritten.
Description: QC database is full.
The oldest result for specified
control and chemistry has been
overwritten. Archiving is
suggested.
• Archive QC so that additional data will
not be overwritten.
0x13A00024
10
QC Database is
full. This result was
not saved.
WLM: QC DB is full.
The result for control:
control name /
Chem: chem name has
not been saved
Description: QC database is full
and there is no space for the result
to be saved. Results may only
overwrite data for existing,
previously defined control and
chemistry information.
• Archive QC data.
0x13A0003F
10
Incomplete set of
calibrator levels in
rack rack ID for the
following
calibrators:
calibrator name list
WLM: Missing
calibrator levels
Rack: rack ID
Calibrators:
calibrator list (File
Name:Line Number)
Description: Not all calibrator
levels were loaded or were loaded
out of order.
• Reload calibrator racks.
• Make sure bar codes are readable.
• If there are 4 calibrators or less, they
must all be in the same rack.
• If more than 4 levels are required,
levels 1–4 must be in the first rack
loaded and the next rack loaded must
contain the remaining levels.
0x13A00056
10
No sample
programming or
pending test found
for samples in rack
#rack ID. Rack will
be unloaded from
the sample
carousel.
WLM: No sample
programming or pending
tests found for sample
on rack. Rack ID: rack
ID (File Name:Line
Number)
Description: A rack was loaded
with nothing to run. The rack is
off-loaded and any previous results
print.
• Check the sample status. It may be
complete and reloaded.
• Could indicate a problem receiving
sample programming from LIS.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action