As-Needed/As-Required Maintenance
Replacing the Sodium Measure/Reference Electrode
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-75
10
6
Using a hemostat, securely clamp Line #22 approximately one inch
(25mm) from the top of the flow cell (1). This will prevent the back flow of
reagent into the flow cell.
NOTICE
Do not disconnect any of the reagent lines attached to the flow cell.
Figure 10.48
7
Disconnect the electrode cable (1).
Figure 10.49
Step
Action
, continued
(2 of 4)