Weekly Maintenance
Clean Probes (Exterior)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-14
October 2005
4
To access the lower portions of the MC and CC sample probes (under the
attached wash collars) remove the sample carousel cover. To continue
access:
• Move the sample probes to a clear space (1) between the sample carousel
racks and the sample carousel tub.
• Press the probe down into the clear area.
• Wipe the outside portion of the probe.
Figure 10.5
5
When all maintenance steps are complete, reinstall covers and select:
• <OK> to clear the pop-up message,
• the Instr Cmd icon from the menu bar,
• <1> Home.
Step
Action
, continued
(3 of 3)