![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 493](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768493.webp)
UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-73
Common Error Messages and Corrective Actions
0x13AD0041
7
Chemistry Version
number on disk
doesn't match
Chemistry Version
number
Backup/Restore:
Chemistry Version
number on disk doesn't
match Chemistry
Version num.
File: file,
Line: line
Description: The contents of the
backup are from a different version
and cannot be restored on this
version.
• Use another disk if available.
1391000E
Cable Down – lost
connection to
Instrument
ROUTER: Cable down
detected. (/mpcvan/src/
cauman/
cau_router.c:1675)
Message ID:
0x1391000E
Description: This Fatal Error will
occur when a network connection
has been lost and the computer
console cannot communicate with
the instrument.
• Make sure that Instrument is on.
• Check Ethernet cable connection at
console, microhub and instrument.
• Reboot system.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action