Monthly Maintenance
Calibrate Lamps and Sensor
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-29
10
Calibrate Lamps and Sensor
Calibrate lamps and/or sensor following monthly cup maintenance, sensor
replacement, or any adjustment made to the module.
18
Select <Close> when priming has completed to exit the pop-up Result
window.
19
Select <Prime5> to prime BUNm/UREAm 5 times with reagent. While
the module is priming:
• Observe the reagent line and reaction cup to confirm there are no leaks.
If any leaks are noticed correct the problem.
• Observe the stir bar to verify that it is rotating properly. Select <Close>
when priming is complete.
20
Select <Drain>. When draining is complete, select <OK>.
21
When priming is complete, select <Close> to exit the Result window and
select <Cancel> to return to the maintenance window.
• If the BUNm/UREAm electrode maintenance was performed, wait 15–
20 minutes to allow for thermal equilibration. Calibrate BUNm/
UREAm. Immediately prior to calibration, wrap the tip of an applicator
stick with lintless tissue, to ensure there are no bubbles in the BUNm/
UREAm cup only. Insert the end of the stick into the bottom of the cup
and carefully swab the electrode.
• The glucose sensor also requires a 15–20 minute equilibration period
prior to performing the sensor calibration.
22
Proceed to
Step
Action
, continued
(5 of 5)
NOTICE
The system will not autoprime in
Maintenance
mode. After completing the
maintenance procedure, or if a procedure is aborted, exit the Maintenance
menu to resume autoprime and return the system to
Standby
.
Step
Action
1
Select the Utils icon from the menu bar.
2
Select <2> Maintenance.
OR
Type 2 in the Option Number field and press [Enter].
(1 of 2)