Four-Month Maintenance
Cleaning MC Reagent Lines
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-54
October 2005
17
CAUTION
Make sure the BUNm/UREAm electrode is slightly withdrawn before
reinstalling the stir bar. This prevents possible damage to the delicate
electrode tip. After the stir bar is reinstalled, tighten the retainer nut
to prevent leakage.
Reinstall the stir bars in the appropriate cups.
Note: Be careful to return the correct size stir bar to its appropriate cup to
assure proper stirring movement. Replace the stir bar if cracks or burrs are
detected. The larger stir bars correspond to the BUNm/UREAm and
GLUCm cups.
18
Select <OK> to continue. Observe whether the stir bars are turning
properly.
CAUTION
The stir bar may rise in the cup due to air accumulation in the lines.
Verify that the stir bar is positioned down into the bottom of the cup.
19
Select <Close> when priming is complete to exit the Result window and
select <Cancel> to return to the Maintenance Menu.
Perform a CUPs Lamp/Sensor Calibration prior to calibrating ALBm,
CREm, GLUCm
a
, PHOSm, and TPm. Refer to Monthly Maintenance,
"
", in this chapter.
a
Only the GLUCm cup module is used on DxC 600 Systems.
Step
Action
, continued
(4 of 4)