Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-68
October 2005
0x13A00092
10
Sample Programs
are too old on rack:
rack number
Position Sample
ID
1 sample ID1
2 sample ID2
3 sample ID3
4 sample ID4
WLM: Sample
programs too old – rack:
rack number [p: sample
ID1] [p: sample ID2]
[p: sample ID3]
[p: sample ID4]
Description: Sample programming
exceeds age specified in setup.
Usually indicates host has reused
sample IDs and the old information
was not cleared from the
instrument.
• Clear the existing sample program and
reprogram or requery host.
0x13A20008
7
MAINT: Cannot
block the run
button.
MAINT: Cannot block
the run button.
File: filename,
line: lineno
Description: Unable to transition to
Maintenance. Samples may still be
running.
• Wait for samples to complete and
system to go to "Standby" before
requesting Maintenance procedure.
0x13A20015
7
MAINT:
ERROR: Motion.
MAINT:
ERROR: Motion.
File: filename,
line: lineno.
Description: Motion error
observed while performing
requested maintenance. Test
aborted, no results are available.
• HOME and restart the maintenance.
0x13A20016
7
MAINT:
ERROR: Reagent
has not been
loaded.
MAINT:
ERROR: Reagent has
not been loaded.
File: filename,
line: lineno.
Description: Maintenance
procedure aborted because the
reagent has not been loaded.
• Load the reagent and retry the request.
0x13A20019
7
MAINT:
ERROR:
Requested rack not
loaded.
MAINT:
ERROR: Requested
rack not loaded.
File: filename,
line: lineno.
Description: Maintenance
procedure aborted because the
requested rack was not loaded.
• The wrong rack was loaded. Check the
rack ID and retry.
• Make sure the bar code faces to the
right when loading.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action