![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 409](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768409.webp)
Event Log
Event Information
UniCel DxC Systems Instructions For Use A13914
Utilities
October 2005
Page 12-9
12
Event Information
Selecting Specific Event Classes
Information Options
After the Event Classes have been selected, the information can be formatted using the
following options.
Table 12.7
Information in the Event Log
Event information
Definition
Number
Sequential number identifying the "order", with number one
representing the most recent entry.
Class
Number corresponding to the 1-10 event classes listed on
the Event Log main screen (e.g., Instrument Events).
Date
The Month, Day, and Year the event occurred.
Time
The time the event occurred in Hours, Minutes, and
Seconds.
Description
Identifies the event or error that occurred.
Step
Action
1
Select the Utils icon from the menu bar.
2
Select <3> Event Log.
OR
Type 3 in the Option Number field and press [Enter].
3
Select the desired Classes. If no Classes are specified, all classes in the
Event Log will be retrieved.
4
Select <Display> to view the status of the desired event.
Table 12.8
Event Classes Information Options
Option
Description
Display
Display Events - for reviewing events on the screen.
Copy
Copy to diskette - for saving events to a disk as a permanent
record, for viewing on another system or sending to Beckman
Coulter.
Time
Date and Time selection - for selecting events for a specified
date and time period.
Clear
Clear Events - for removing unwanted event information. To
clear information, select the event class(es) to be deleted and
select the clear confirmation window.
(1 of 2)