CTS Wick Replacement Procedure
System Preparation
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-122
October 2005
System Preparation
The Cap Piercer Blade must be lowered before the Wick is replaced. Perform the
following steps:
Remove the Cap Piercer Cover
The Cap Piercer Cover must be removed to get access to the Cap Piercer Assembly.
Refer to
and perform the following steps:
Step
Action
1
Select the Utils icon from Main menu bar.
2
Select <2> Maintenance.
3
Select <1> 1-Blade Thick CTS Blade Replacement.
4
Select <Continue>. Wait for the Maintenance Confirmation screen.
CAUTION
DO NOT select <OK> on the screen when the cover is removed. This
would move the Cap Piercer upward and cause a pinching hazard.
Step
Action
1
Lift the instrument cover.
2
Use a flat-edge screwdriver to fully loosen the two Screws (1) from Cap
Piercer Cover (2).
Figure 10.81
3
Lift and remove the Cap Piercer Cover and set it aside.
A007556P.EPS
1
2