Control ID Assignments
Introduction to Assigning QC Bar Codes
Quality Control
UniCel DxC Systems Instructions For Use A13914
Page 6-8
October 2005
Control ID Assignments
Introduction to Assigning QC Bar Codes
A maximum of eight bar codes (control IDs) per control may be defined. Even if a
predefined control ID exists, before running the test, first use the sample programming
screen to program the tests for the bar-coded controls. Assignment of bar code IDs is
mandatory for the system to process duplicate controls.
Assigning QC Bar Codes
Follow the steps below to assign QC bar codes.
Printing Bar Code Assignments
Press the [Print Screen] key to print the Control ID assignment window.
Step
Action
1
From the Define/Edit Screen, select Control ID [F2].
2
The Control ID dialog box comes into view. A maximum of eight unique
bar code IDs with a maximum of 15 alphanumeric characters may be
entered. Press the [Tab] key after each bar code entry to move between
fields.
When creating Control Bar Code IDs, use a format that is distinctly
different from Sample IDs. This will prevent the reporting of erroneous
results due to Controls being run as patient samples, or patient samples
being run as controls.
Example:
Control Bar Code ID: SYNCHRON01
Sample Bar Code ID: 0000001
3
Select <OK> to exit the screen and save the bar code IDs.
OR
Select <Cancel> to clear all modifications.