Defining a Control
Defining a Control
Quality Control
UniCel DxC Systems Instructions For Use A13914
Page 6-6
October 2005
8
Select Chems [F1]. A list of configured chemistries will appear. Add or
remove chemistries by selecting and adding to the list, or by selecting and
removing from the list. Use the scroll bar to access all chemistries.
Select <OK> to leave the selection screen and save the chemistry
selections.
Select <Cancel> to exit without changing the chemistry selection.
9
Type the following information in the appropriate fields for all chemistries
shown:
Field
Description
QC File Number
Type a QC file number for each chemistry.
• The QC file number must be a unique
3-digit number from 1 to 999.
• If a QC file number has already been
assigned, the system will not accept it a
second time. Refer to "
" in this chapter to obtain a list of used
QC File numbers.
Assigned Mean
Type the assigned mean for the chemistry
level and lot. The system adds zeroes to the
end of the number to show maximum
precision (X.XXX).
Assigned SD
Type the assigned standard deviation, ±1SD.
The system adds zeroes to the end of the
numbers to show maximum precision
(X.XXX).
Constituent Code
This field is optional. If desired, type the
code constituent code.
Step
Action
, continued
(2 of 3)