Two-Month Maintenance
Replace CTS Blade/Wick (CTS Option only)
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-41
10
7
Refer to
Hold the blade holder (5) with the blade clamp hole (3) on the top as shown
below. Note that the blade clamp hole (3) is keyed and the blade clamp (2)
has a slot. Install the blade (1) into the blade holder (5) as follows:
• Put the blade clamp end (2) of the blade into the blade clamp hole (3).
• Rotate the blade, if necessary, until it falls into place.
• Finger tighten the captive blade screw (4).
• Remove the protection from the points (6) of the blade.
• Do not touch the points of the blade.
Figure 10.24
8
Refer to
Carefully move the points of the blade down through the wash tower and
through the alignment slot at the bottom:
• Push down on the blade holder screws (1) and align them over their
holes.
• Slowly lower the blade through the alignment slot until the blade holder
screws are in their holes. Be careful not to hit the points of the blade on
the edge of the metal alignment slot.
• The blade is through the alignment slot when the blade holder touches
the surface of the frame of the cap piercer.
• Secure clips to blade holder wash block.
• Install and tighten the two captive blade holder screws (1). Tighten the
captive blade screw (2).
Step
Action
, continued
(4 of 5)
A015941P.EPS
5
6
1
3
4
2