Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-32
October 2005
Figure 10.16 CC Sample and Reagent Mixers
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
Refer to
. Replace the Cap Piercer blade approximately once a month. If
the system runs >500 samples per day, replace the blade more frequently. The blade
count records the number of caps pierced by the blade. Refer to
Figure 10.17 Cap Piercer Blade
1. CC Sample Mixer
2. CC Reagent Mixer
A015939P.EPS
2
1
CAUTION
The Cap Piercer blade is very sharp and has been exposed to potentially
biohazardous fluids. To prevent injury or exposure, do not touch the points
of the blade and wear gloves.