Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-34
October 2005
0x01FA001A
4
10 PSI Air Supply
Pressure Is High.
CAU: 10 PSI Air
Pressure Is High.
Message ID:
0x01FA001A
Description: This error occurs
when the transducer line #129 is
outputting a value into the Hydro
Board that is above the 10 psi
specification.
• Check pressure levels and HOME the
system.
• Adjust pressure levels if necessary.
First adjust pressure in Main system
above 17 psi. Then adjust the high
pressure system to 17 psi. Finally,
adjust the low pressure system. The
pressure range for the low pressure
system is 8–11 psi. Set it at 10 psi.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x01FA001C
4
17 psi air supply
pressure is high.
CAU: 17 psi air pressure
is high
Description: This error will occur
whenever there is a problem within
the high-pressure system. If the
High pressure reads higher than
19 psi for more than 15 seconds it
will be flagged as "High."
• Check pressure levels and HOME the
system.
• Adjust pressure levels if necessary.
First adjust pressure in Main system
above 17 psi. Then adjust the high
pressure system to 17 psi.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x01FA0034
4
Hydro Illegal
Switch Condition.
CAU: Hydro Illegal
Switch Condition
Message ID:
0x01FA0034
Container: container
Description: Float Switch indicate
container is full and empty at the
sample time.
• Unscrew the canister. Wipe the float
switch and make sure the float isn't
stuck.
• Replace the Float Switch in the
specified container.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action