Prime
Priming MC Subsystems
Utilities
UniCel DxC Systems Instructions For Use A13914
Page 12-6
October 2005
Priming MC Subsystems
MC Subsystem Prime allows the priming of the cup modules, the ISE flow cell and
EIC, and the Sample Delivery Subsystem.
• If a cup selection is made and neither reagent or DI water is specified, only DI water
will be primed into the cup.
• If both reagent and DI water are selected, both will be primed and the cup(s) will be
left full of DI water.
• Whenever Acid is primed, Reference must also be primed to prevent acid shocking
the electrodes. Therefore there is no Acid-only option.
• Only one ISE selection can be made at a time. However the ISE All option can be
used to select the entire list.
Follow the steps below to prime the MC Subsystems.
Step
Action
1
From the Utilities dialog box, select <1> Prime.
OR
Type 1 in the Option Number field and press [Enter].
2
Select MC [F4].
3
Select the desired cup(s) and specify Reagent and/or DI water.
AND/OR
Select a single ISE option by choosing a check box.
AND/OR
Select the Sample Delivery Subsystem check box.
4
Type the desired number of primes (1–99) in the Number of primes to
repeat field.
5
Select <Start Prime>. The system status changes to Priming and the
number of primes remaining will be shown.
6
Upon completion of the requested prime, the number of primes remaining
will be blank and the system status returns to Standby.