![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 392](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768392.webp)
Shutdown
Introduction
System Status and Commands
UniCel DxC Systems Instructions For Use A13914
Page 11-16
October 2005
Shutdown
Introduction
Shutdown is used to place the system into a state where it is safe to reset or turn power
off. System files are closed to prevent data corruption.
Additional Information
For additional information about preparing the system to remain in a powered-off
condition for an extended period, refer to the Extended Shutdown in the UniCel DxC
Synchron Clinical Systems Reference Manual.
How to Shutdown
NOTICE
When performing a shutdown, it is very important to read and follow the
instructions on the screens. Do not turn the console off until "
Shutdown:
System may now be powered down
" appears on the monitor. If power is
turned off before this, there is a potential for data corruption.
Step
Action
1
Verify that the system is in Standby or Stopped.
2
Select the Instr Cmd icon from the menu bar.
3
Select <4> Shutdown.
OR
Type 4 in the Command No. field and press [Enter].
4
Select <OK> or <Cancel>.
If <OK> is selected, the Shutdown in Progress screen appears with the
message:
"Please wait for further instructions. DO NOT TURN THE POWER OFF."
The system status goes to Shutdown.
5
From the Select Reboot or Shutdown dialog box, select the appropriate
option:
<Reboot> to return the system to an operational state. The system will
automatically reboot and return to the Main screen.
OR
<Shutdown> to prepare the system to have the power turned off. A
counter appears while the system shuts down. When "Shutdown: System
may now be powered down" appears in the top, left corner of the screen,
then it is safe to turn off the console power.