As-Needed/As-Required Maintenance
Cleaning CC Sample Probe, Reagent Probes and Mixers
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-108
October 2005
Cleaning CC Sample Probe, Reagent Probes and Mixers
The following procedure describes the automated procedure to wash the CC sample
and reagent probes and mixers. This routine will take the system approximately 15
minutes.
Material Required:
• CCWA cartridge P/N 657133
NOTICE
The system will not autoprime in
Maintenance
mode. After completing a
maintenance procedure, or when a procedure is aborted, exit the Maintenance
menu to resume autoprime and return the system to
Standby
.
Step
Action
1
Load the CCWA cartridge onto the system by selecting:
• the Rgts/Cal icon from the menu bar,
• An empty chem position to load (CC side),
• Load [F1]. Scan Chemistry Wash Solution Cartridge (CCWA).
2
The Wash procedure is accessed by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <2> Cartridge Chemistry Probe Cleaning.
OR
• Type 2 in the Procedure No. field and press [Enter].
Follow all instructions listed on the screen.
3
Select <Cancel> to exit this screen without performing the procedure.
OR
To perform the procedure, select <Continue>.
NOTICE
If desired, you may exit this procedure at any time before it is completed
by selecting
STOP
on the instrument; wait for the system to finish
processing. To return the system to
Standby
, select
Instr Cmd
from the
menu bar, then select
Home
.
A pop-up message will appear to confirm that probe and mixer cleaning is
complete.
Select <Close> to exit.
4
At the completion of this procedure, the Maintenance menu will appear.
(1 of 2)