Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-70
October 2005
0x13A70005
7
Instrument not in
proper state.
ICMD: Can't do HOME
in current instrument
state.
File: filename,
line: line number
Description: System is not in
proper state for Homing (usually
still Stopping). Must be Standby/
Stopped/Disabled/PVT/Diag/Maint
or Align.
• Wait for the system to get to Standby/
Stopped/Disabled/PVT/Diag/Maint or
Align.
• Reboot if necessary.
0x13A70006
7
Instrument not in
proper state.
ICMD: Can't do
PAUSE in current
instrument state.
File: filename,
line: line number
Description: The instrument is not
in the proper state to accept a
PAUSE command. Only select
PAUSE when not already in a
pausing state and at least one side
of the instrument is running.
• If Pause was previously selected, the
System will go to "Standby"
automatically.
• Select Pause only when at least one
side of the system is running.
• Reboot if necessary.
0x13A70007
7
Instrument not in
proper state.
ICMD: Can't do
SHUTDOWN in current
instrument state.
File: filename,
line: line number
Description: System is not in
proper state for Shutdown
command. Only select Shutdown
when all sides of the system are in
Standby/Stopped/or Disabled
states. Press the STOP button if
necessary to obtain these states.
• Wait for the system to reach "Standby"
or "Stopped".
• To immediately shutdown, press the
STOP button and wait to reach
"Stopped", then request Shutdown.
0x13AC0008
7
Cannot create
socket.
BOOT: Cannot create
socket
Description: Unable to create a
network socket (shutdown didn't
complete cleanly).
• Reboot.
0x13AC000B
7
Select failed.
BOOT: Select failed
Description: Unable to boot.
• Make sure that the instrument power is
on and Reboot. Continued errors may
indicate an incorrect hardware address.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description