CTS Wick Replacement Procedure
Clean the Blade
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-127
10
2
Refer to
. Wipe the Blade with a cotton-tip applicator as
shown.
CAUTION
The points on the end of the Blade are very sharp and have been
exposed to potentially biohazardous fluids. To prevent possible injury
or exposure, avoid the Blade area.
Figure 10.86
3
Put the Blade Assembly back on the same tray.