Priming Hydropneumatic Subsystems
UniCel DxC Systems Instructions For Use A13914
Utilities
October 2005
Page 12-5
12
Priming Hydropneumatic Subsystems
The Hydropneumatics Subsystem allows independent priming of each of its
components. Follow the steps below to prime any combination of the components.
Step
Action
1
From the Utilities dialog box, select <1> Prime.
OR
Type 1 in the Option Number field and press [Enter].
2
Select CC [F2].
3
Select one or a combination of the following check boxes:
• Reagent Delivery Subsystem
• Sample Delivery Subsystem
• Cuvette Wash.
4
Type the desired number of primes (1–99) in the Number of primes to
repeat field.
5
Select <Start Prime>. The system status changes to Priming and the
number of primes remaining are shown.
6
Upon completion of the requested prime, the number of primes remaining
will be blank and the system status returns to Standby.
Step
Action
1
From the Utilities dialog box, select <1> Prime.
OR
Type 1 in the Option Number field and press [Enter].
2
Select Hydro [F3].
3
Select any or all of the hydropneumatic subsystems.
4
Select <Start Prime>. Although the instrument status remains in Standby,
the hydro prime will begin.