As-Needed/As-Required Maintenance
Replacing the CO
2
Membrane
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-80
October 2005
5
Locate the CO
2
measuring electrode. Place absorbent tissue beneath the
CO
2
electrode port. Using hemostats, clamp off reagent Lines #33 (1) and
#32 (2) as closely as possible to the CO
2
electrode. (Line #32 connects as a
loop to the CO
2
reference electrode.) Remove electrode by turning the
electrode retainer nut counterclockwise.
Figure 10.54
Step
Action
, continued
(2 of 3)