As-Needed/As-Required Maintenance
EIC Port Cleaning
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-89
10
7
Hold the removed EIC over a sink or absorbent tissue. Using a 20 or 30 mL
irrigation syringe filled with 10% bleach solution, insert the tip of the
syringe into the valve port opening (1).
Figure 10.66
8
Flush the 10% bleach solution through the valve port. This flushes the
sample inject area. Allow the bleach solution to sit for 1 minute.
Note: The 1-minute waiting period applies only to the 10% bleach solution,
and not to the following deionized water or 70% isopropyl alcohol.
9
Repeat the flushing procedure through the valve port using deionized
water.
10
Repeat the flushing procedure through the valve port using 70% isopropyl
alcohol.
11
Repeat the flushing procedure through the valve port using deionized
water.
Step
Action
, continued
(4 of 7)