As-Needed/As-Required Maintenance
Decontaminating the Instrument
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-119
10
2
Open the lower, middle compartment door; release latch (1) on the
hydropneumatics drawer and pull out the hydropneumatic system.
Figure 10.79
3
Remove the Wash Concentrate bottle (1), the DI Water Reservoir canister
(2), and the DI Water canister (3) from the system. If necessary, use the
spanner wrench tool to remove the canisters.
Figure 10.80
Step
Action
, continued
(2 of 3)
A016466P.EPS
1
A016489P.EPS
1
2
3