As-Needed/As-Required Maintenance
EIC Port Cleaning
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-87
10
3
Disconnect the solenoid valves, located on each side of the EIC, at the
valve connector (1) as shown below on DxC 800 Systems. The connector
is located near the EIC ports where the ends of the two solenoid valve
cables meet. For DxC 600 Systems, trace the solenoid wiring to its source
and disconnect. Pull apart the valve connector to disconnect.
Figure 10.62
4
Carefully remove the Phillips screw and washer (1) (screw removed in
figure) located on the top of the EIC. Lift off the EIC (two sections with
solenoid valves (2) attached) from the drip tray (3).
NOTICE
DO NOT REMOVE the two screws on the drip tray. Removal or
repositioning of the drip tray requires an MC sample probe alignment.
Figure 10.63
Step
Action
, continued
(2 of 7)