![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 469](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768469.webp)
UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-49
Common Error Messages and Corrective Actions
0x03880006
3
CC Reaction
Carousel
Temperature Out of
Range
CAU: cc rwHtr reaction
wheel temp out of range
Message ID:
0x03880006
Subsystem: cc rwHtr
Description: This error occurs
under the following conditions:
• When the reaction wheel monitor
sensor readings fall outside of the
+37 ± 1.0ºC specification.
• When the reaction wheel control
sensor readings fall outside of the
+37 ± 0.1ºC specification.
• Check System Status for Reaction
Carousel Temperature.
• Check Cuvette Status for a cuvette
flagged in red. This may indicate
broken or cracked cuvette.
• If a cuvette appears to be damaged,
replace it (extra cuvettes available in
Spares kit and may be ordered).
• Press the STOP button and HOME the
system.
0x04B2001C
4
MC Sample Level
Sense Error
Sample ID: sample
id
Rack: rack id
Position: rack
position
CAU: Sample Level
Sense Error
Message ID:
0x04B20001C
Sample ID: sample id
Rack: rack id
Description: This error occurs
under following conditions:
• The bottom of the container is
reached before sample is detected
(e.g., empty cup or tube).
• Insufficient sample volume in
container.
• Level sense is triggered sooner
than expected for the given
container type.
• Check sample for adequate volume.
• Check proper rack assignment for type
of container used (reserved vs.
non-reserved).
• Look for fluid in collar wash – fluid in
collar wash can trigger false level
sense.
• Check alignment of sample probe in
collar wash.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action