As-Needed/As-Required Maintenance
EIC Port Cleaning
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-88
October 2005
5
Place absorbent tissue underneath the EIC port area. Place clamps or
hemostats on lines #18 (1), #24 (2), #23 (3), and #26 (4), near the EIC
ports. Remove EIC lines from all the ports #18, #24, #23, #26, and #15 (5).
Figure 10.64
6
Carefully loosen the two screws (1) from the solenoid valve located on the
back side of the EIC. Remove the solenoid valve and set aside.
Figure 10.65
Step
Action
, continued
(3 of 7)