Shutdown
On/Off Switch
UniCel DxC Systems Instructions For Use A13914
System Status and Commands
October 2005
Page 11-17
11
On/Off Switch
The circuit breaker/power ON/OFF switch is located behind the front right-hand door
of the instrument. (Refer to
.)
Figure 11.1
ON/OFF Switch
Line Converter Device
All power to the instrument is passed through the line converter device located in the
lower right compartment of the instrument. The line converter serves as a conditioner
to limit the effect of line voltage changes that could damage the sensitive electronic
components of the system.
DC Output Power Supplies
Power is routed to redundant DC output power supplies which provide 24V or 36V
power for distribution throughout the system. By using redundant power supplies, not
all are used at any one particular time. Should one fail, the operator is alerted, but may
continue operation unaffected.
Uninterruptible Power Supply (UPS)
The computer console requires an additional power source. The UPS (Uninterruptible
Power Supply) plugs into a standard 110–120V AC, 50/60 Hz power source. The UPS
in turn supplies power to the computer console and the monitor as well as providing a
short-term battery backup in case of a power failure. The printer plugs directly into its
own standard (110–120V AC, 50/60 Hz) power source.