![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 481](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768481.webp)
UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-61
Common Error Messages and Corrective Actions
0x05DC0227
4
No Response From
Hydro Subsystem
CAU: Homing Timeout
for Hydro Subsystem.
Message ID:
0x05DC0227,
Subsystem: boot,
homing command
Description: This error occurs
when the hydro fails to respond to
commands or fails to attach.
This event can be generated by any
Hydro command sent by start.dat
including "AttachHydro",
StartAutomonitor"
"HydroInitStatus", or
"HydroWatchdogOn".
• Check Event Log for specific
subsystem errors.
• Press the STOP button and HOME the
system.
• Reboot.
• If the error continues, contact Beckman
Coulter Technical Support Center for
additional information.
0x13940013
7
PVT:
ERROR:
Insufficient
reagent.
PVT:
ERROR: Insufficient
reagent.
File: filename,
line: lineno.
Description: Cannot perform the
test because there is not enough
reagent.
• Load a new PVT reagent cartridge.
0x13940014
7
PVT:
ERROR:
Insufficient sample.
PVT:
ERROR: Insufficient
sample.
File: filename,
line: lineno.
Description: Sample Level Sense
error while running PVT.
• Pour a fresh PVT sample and restart
the test.
0x13940015
7
PVT:
ERROR: Motion.
PVT:
ERROR: Motion.
File: filename,
line: lineno.
Description: A motion error
occurred while running a PVT. The
test is aborted and no result will be
shown.
• HOME and restart the test.
0x1394001B
7
MAINT: Cannot
block the run
button.
PVT: cannot block the
run button.
File: filename,
line: lineno.
Description: Unable to transition to
Maintenance. The system may still
be in a running state.
• Allow samples to complete before
requesting maintenance.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action