Home is used to move mechanical assemblies to a known "home" position and primes
the system.
Why Use Home
Home is useful in the following situations:
• to attempt recovery from a motion error.
• to return the system or one component of the instrument from a Stopped state to
Standby.
How to Home
Note: If one side of the instrument (MC or CC) is Stopped and the other side is
running, the Home command operates on the side that is Stopped. The other side of the
instrument continues to run.
NOTICE
If the
STOP
button is pressed immediately after selecting
Home
:
• Wait at least 10 seconds.
• Press the
STOP
button again.
This makes sure that the system correctly reaches the
Stopped
state.
Step
Action
1
Select the Instr Cmd icon from the menu bar.
2
Select <1> Home.
3
The instrument status will change to Homing. If the home completes
successfully, the instrument will be in Standby. Should an error occur, the
status will be Stopped and the appropriate error message appears.