Processing Parameters
Third Inject: Component
UniCel DxC Systems Instructions For Use A13914
User Defined Reagents
October 2005
Page 9-13
9
Third Inject: Component
This designates the compartment of the cartridge which will be pipetted third. If Third
Inject is not used, the compartment selection is None and should not be modified.
Third Inject: Dispense Volume
The combined volume of all inject reagents must be between 200 and 327 L.
Third Inject: Add Time
Add Time programs the reagent dispense time into the cuvette.
• The Third Inject may be added a minimum of 8 seconds after reagent(s) and/or
sample have been pipetted. (Time is entered in increments of 1 second; however,
the reagent will be added at the 8-second interval closest to the time defined.
Sample addition occurs at 1 second.)
• A positive number indicates Third Inject reagent component is added after sample
add.
• Third Inject add time must be greater than Second Inject add time.
Sample Volume
Total volume of sample and reagents A, B, and C cannot exceed 330 µL. Volume is
specified in increments of 1 µL.
Table 9.19
Options for the Third Inject Component
None, A, B, or C
Table 9.20
Options for the Third Inject Dispense Volume
6 to 75 L
Table 9.21
Allowable Entries for Add Time for the Third Inject
-172, 9 to 738 seconds
Table 9.22
Allowable Entries for Sample Volume
3 to 40 µL