CTS Wick Replacement Procedure
Introduction
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-121
10
CTS Wick Replacement Procedure
Introduction
This procedure applies to the UniCel DxC 600/800 and the UniCel DxC 600i Systems
with the Closed Tube Sampling (CTS), 1-Blade Thick Option. The wick is found
within the Cap Piercer Assembly.
The Cap Piercer wick and blade are normally replaced every two months during
scheduled maintenance. However, if blood is not continuously wiped from the sample
tube caps before they are loaded onto the instrument, the CTS blade/wick area will
become dirty. This may cause a contaminated condition. As a result, use this procedure
to replace the wick as needed.
Materials Required
Collect the following materials:
• 1 medium flat-edge screwdriver
• 1 medium Phillips screwdriver (for DxC 600i systems)
• 1 small tweezers
• 2 plastic trays (approximately 9 inches × 12 inches) (23.9 centimeters × 30.5
centimeters)
• Absorbant paper towel
• 1 bottle of Auto-Gloss
• 1 Wick Kit (P/N A26924)
• Cotton-tip applicators (6 inches [15.2 centimeters] long, or greater)
• Disposable pipettes
• Personal protective gear such as rubber gloves and eye shields
• Scissors
CAUTION
All biohazard precautions should be observed when doing maintenance,
service, or troubleshooting activities on the system. This includes but may
not be limited to wearing gloves and eye shields, and washing hands after
working on contaminated portions of the system.