leshooting Calibr
ation and Result Errors
October 2005
13-53
Common Error Messages and Corrective Actions
0x04B30013
4
MC Sample Probe
disabled due to
obstruction errors
CAU: MC Sample
Probe disabled due to
obstruction errors
Message ID:
0x04B30013
Description: More than 3 samples
in a row had obstruction errors
(clots), MC side "Stopped".
• Prime MC internal probe wash to make
sure it is not obstructed.
• Probe may need to be replaced.
• If error continues, perform MC ODC
Calibration (Alignment).
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action