Error Code – Definitions
Less Than X-ORR LO
UniCel DxC Systems Instructions For Use A13914
Troubleshooting Calibration and Result Errors
October 2005
Page 13-25
13
Less Than X-ORR LO
Calculated result is less than the lower Reportable Range. Result is not reported.
Level Sense Error in Diluent Cartridge
A Level Sense Error occurred in the diluent cartridge and the system was unable to
perform the required dilutions. This error may affect IGG, IGM, IGA, and TRF in
serum samples or MG, BUN, UREA, or URIC in urine mode. All of these tests
perform on-board dilutions before analysis.
Low Reagent Fluid
A Level Sense Error was obtained when attempting to run a specific cartridge.
Remove cartridge and check for bubbles and adequate reagent level. Remove bubbles,
if present, and reload. If system still receives a level sense error, refer to the UniCel
DxC Synchron Clinical Systems Reference Manual, CHAPTER 5, Advanced
Troubleshooting, CC Hardware Symptoms for additional investigation.
Low Sample Fluid
No sample detect.
MC Reagent Level High/Low
Reagent level detected with cup is either too high or too low.
A high flag may indicate issues with a blocked drain value, low vacuum or
malfunctioning level sensing. A low flag may indicate an empty reagent bottle or
loose straw, bubbles in the reagent, pump failure, disconnected reagent lines or valve
failures. After the problem is solved, repeat the last two samples run in that cup
module.
Motion Error
Motion error occurred while system was performing requested test.
No Diluent on Reagent Carousel
The required diluent cartridge is not present on the system and the requested tests
cannot be performed.
Not Ready
The system is preparing to run the requested tests and the tests are Pending.
ORDAC Result
Result was calculated using ORDAC feature.