Operating Manual 1164.4556.12 - 06
4.21
R&S FSP
Instrument Functions
Analyzer Mode
In time domain (span = 0 Hz), the range of sweep times is 1 µs to 16000 s is select-
able in steps of max. 5% of the sweep time. For numeric inputs, rounding is made to
the nearest possible sweep time. For rotary knob or UP/DOWN key inputs, the
sweep time is adjusted in steps either downwards or upwards.
The manual input mode of the sweep time is indicated by a green asterisk (*) on the
display. If the selected sweep time is too short for the selected bandwidth and span,
level measurement errors will occur. This happens because the available settling
time for the resolution or video filters is too short. In this case, the R&S FSP outputs
UNCAL
on the display and marks the indicated sweep time with a red asterisk (*).
Remote command:
SWE:TIME:AUTO OFF;
SWE:TIME 10s
RES BW AUTO
The
RES BW AUTO
softkey couples the resolution bandwidth to the selected span.
Changing the span causes automatic adjustment of the resolution bandwidth.
Automatic coupling of resolution bandwidth to span is always recommended when a
favorable setting of the resolution bandwidth in relation to the selected span is
desired for the measurement under request.
The coupling ratio is set in the
COUPLING RATIO
submenu.
The
RES BW AUTO
softkey is only available in the frequency domain (span > 0 Hz).
The softkey is deactive in the time domain.
Remote command:
BAND:AUTO ON
VIDEO BW
AUTO
The
VIDEO BW AUTO
softkey couples the video bandwidth to the resolution band-
width. If the resolution bandwidth is changed, the video bandwidth is automatically
adjusted.
The coupling of the video bandwidth is always recommended when the minimum
sweep time is required for a selected resolution bandwidth. Narrower video band-
widths require longer sweep times due to the longer settling time. Wider bandwidths
reduce the signal/noise ratio.
The coupling ratio is set in the
COUPLING RATIO
submenu.
The coupling of the video bandwidth to the resolution filter is also permitted for the
time domain display (span = 0).
Remote command:
BAND:VID:AUTO ON
SWEEPTIME
AUTO
The
SWEEPTIME AUTO
softkey couples the sweep time to the span, video band-
width (VBW) and resolution bandwidth (RBW). The sweep time is automatically
adjusted on any change in span, resolution bandwidth or video bandwidth.
The softkey is only available in the frequency domain (span >0 Hz). It is blanked in
the time domain.
The R&S FSP always selects the shortest sweep time possible without falsifying the
signal. The maximum level error compared to using a longer sweep time is < 0.1 dB.
If additional bandwidth and level errors are to be avoided, the sweep time is to be
set to three times the time offered in coupled mode.
The frequency scan rates required when using the quasipeak detector are defined in
Annex B in CISPR 16-2-x:
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...