![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 628](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864628.webp)
R&S FSP
Remote Control – Description of Commands
TRACe Subsystem
6.234
Operating Manual 1164.4556.12 - 06
6.20 TRACe Subsystem
The TRACe subsystem controls access to the instrument's internal trace memory.
6.20.1
General Trace Commands
TRACe<1|2>:COPY
TRACE1| TRACE2| TRACE3 , TRACE1| TRACE2| TRACE3
This command copies data from one trace to another. The second operand describes the source, the
first operand the destination of the data to be copied.
The numeric suffix at TRACe<1|2> selects the measurement window.
Example:
"TRAC:COPY TRACE3,TRACE1"
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
This command is an event and therefore has no query and no *RST value.
TRACe<1|2>[:DATA]
TRACE1 | TRACE2 | TRACE3 | LIST | SPURious | ABITstream | PWCDp |
CTABle, <block> | <numeric_value>
This command transfers trace data from the control computer to the instrument, the query reads trace
data out of the instrument.
The numeric suffix at TRACe<1|2> selects the measurement window.
Parameter:
TRACE1
to
TRACE3
selects trace 1 to 3.
LIST
reads the peak list in the spurious measurement
(for
details on this measurement see page 220).
The suffix at TRACe<1|2> is irrelevant.
As results a list of <result of range 1>,< result of range 2>,..< result of range n> are
returned.
Every single range has following format:
<No>,<Start>,<Stop>,<rbw>,<freq>,<Levelabs>,<Levelrel>,<Delta>,
<Limitcheck>,<unused1>,<unused2>
Where:
No
Range number
Start
Range start frequency
Stop
Range stop frequency
Rbw
Resolution bandwidth
Freq
Frequency of the peak in the range
Levelabs
Absolute peak power of the range in dBm
Levelrel
Reserved (0.0)
Delta
Delta of the peak power to the limit line in dB
Limitcheck
Limit check state (0 = PASSED, 1 = FAILED)
Unused1
Reserved (0.0)
Unused2
Reserved (0.0)
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...