![Rohde & Schwarz R&S FSP Series Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864087.webp)
Operating Manual 1164.4556.12 - 06
4.15
R&S FSP
Instrument Functions
Analyzer Mode
REF LEVEL
The
REF LEVEL
softkey allows the reference level to be input in the currently active
unit (dBm, dBµV, etc.)
Remote command:
DISP:WIND:TRAC:Y:RLEV -60dBm
RANGE LOG
100 dB
The
RANGE LOG 100 dB
softkey sets the level display range to 100 dB.
Remote command:
DISP:WIND:TRAC:Y:SPAC LOG
DISP:WIND:TRAC:Y 100DB
RANGE LOG
MANUAL
The
RANGE LOG MANUAL
softkey activates the manual entry of the level display
range. Display ranges from 1 to 200 dB are allowed in 10 dB steps. Inputs which are
not allowed are rounded to the next valid value.
The default setting is 100 dB.
Remote command:
DISP:WIND:TRAC:Y:SPAC LOG
DISP:WIND:TRAC:Y 120DB
RANGE
LINEAR
The
RANGE LINEAR
softkey selects linear scaling for the level display range of the
R&S FSPr. In addition, it opens a submenu for selecting % or dB for the scaling.
When linear scaling is selected, the % scaling is first activated (see also
RANGE
LINEAR dB
softkey).
Remote command:
DISP:WIND:TRAC:Y:SPAC LIN
RANGE
LINEAR %
The
RANGE LINEAR %
softkey selects linear scaling in % for the level display
range, i.e. the horizontal lines are labelled in %. The grid is divided in decadic steps.
Markers are displayed in the selected unit; delta markers are displayed in % refer-
enced to the voltage value at the position of marker 1.
Remote command:
DISP:WIND:TRAC:Y:SPAC LIN
RANGE
LINEAR dB
The
RANGE LINEAR dB
softkey selects linear scaling in dB for the level display
range, i.e. the horizontal lines are labelled in dB.
Markers are displayed in the selected unit; delta markers are displayed in dB refer-
enced to the power value at the position of marker 1.
Remote command:
DISP:WIND:TRAC:Y:SPAC LDB
!
Side menu
(option B25)
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...