
R&S FSP
Instrument Functions
External Generator Control – Option R&S FSP-B10
4.234
Operating Manual 1164.4556.12 - 06
4.9.3.1
Calibration of Reflection Measurement
The calibration mechanism essentially corresponds to that of the transmission mea-
surement.
CAL REFL
SHORT
The
CAL REFL SHORT
softkey starts the short-circuit calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration curve is
formed by averaging the two measurements and stored in the memory. The order of
measurements is optional.
After the calibration the following message is displayed:
The display is cleared after approx. 3 seconds.
Remote command:
CORR:METH REFL
CORR:COLL THR
CAL REFL
OPEN
The
CAL REFL OPEN
softkey starts the open-circuit calibration. During calibration
the following message is displayed:
Remote command:
CORR:METH REFL
CORR:COLL OPEN
4.9.4
Calibration Mechanism
Calibration means a calculation of the difference between the currently measured
power and a reference curve, independent of the selected type of measurement
(transmission/reflection). The hardware settings used for measuring the reference
curve are included in the reference data set.
Even with normalization switched on, the device settings can be changed in a wide
area without stopping the normalization. This reduces the necessity to carry out a
new normalization to a minimum.
For this purpose, the reference dataset (trace with 625 measured values) is stored
as a table with 625 points (frequency/level).
Differences in level settings between the reference curve and the current device set-
tings are taken into account automatically. If the span is reduced, a linear interpola-
tion of the intermediate values is applied. If the span increases, the values at the left
or right border of the reference data set are extrapolated to the current start or stop
frequency, i.e. the reference data set is extended by constant values.
An enhancement label is used to mark the different levels of measurement accu-
racy. This enhancement label is displayed at the right diagram border when normal-
ization is switched on and a deviation from the reference setting occurs. Three
accuracy levels are defined:
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...