R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.44
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:LIMit<1...8>:UPPer:THReshold
<numeric_value>
This command defines an absolute threshold value for limit lines with relative y-axis scaling
independently of the measurement window. The absolute threshold value is used in the limit check as
soon as it exceeds the relative limit value.
The unit must correspond to the unit selected with
CALC:LIM:UNIT
(except dB which is not possible).
If no unit is indicated, the unit defined with
CALC:LIM:UNIT
is automatically used (exception: dBm
instead of dB).
In analyzer mode, the units DEG, RAD, S, HZ, PCT are not available.
Example:
"CALC:LIM2:UPP:THR -35DBM"
'Defines an absolute threshold value for limit line 2.
Characteristics:
*RST value: -200 dBm
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...