Operating Manual 1164.4556.12 - 06
4.123
R&S FSP
Instrument Functions
Analyzer Mode
The frequency range where measurements are actually performed is set by the start
and stop frequency parameters of the R&S FSP; these parameters are independent
of the sweep ranges. It is thus possible to define sweep ranges for a measurement
task that can be stored and reloaded and to quickly and easily set the frequency
range to be actually measured by means of two parameters; complex editing in the
sweep table is not necessary.
SPURIOUS
(ON OFF)
The
SPURIOUS (ON OFF)
softkey switches the spurious emissions measurement
on or off according to the current configuration.
Remote command:
SWEEP:MODE LIST
switches the spurious list on
SWEEP:MODE AUTO
switches the spurious list off
SWEEP LIST
The
SWEEP LIST
softkey opens a submenu where predefined sweep ranges can
be edited, or new ranges generated or deleted. A table listing the current sweep
ranges is displayed.
Remote command:
--
EDIT SWEEP
LIST
The
EDIT SWEEP LIST
softkey opens the table for editing the sweep ranges.
When a limit line is defined in steps, the weaker limit is used at the frequency point
with the straight vertical section.
SWEEP LIST
RANGE 1
RANGE 2
RANGE 3
RANGE 4 RANGE 5
Range Start
9 kHz
150 MHz
30 MHz
Range Stop
150 kHz
30 MHz
1 GHz
Filter Type
NORMAL (3dB) NORMAL (3dB) NORMAL (3dB)
RBW
1 kHz
10 kHz
100 kHz
VBW
3 kHz
30 kHz
300 MHz
Sweep time mode
AUTO
AUTO
AUTO
Sweep time
145 ms
300 ms
100 ms
Detector
RMS
RMS
RMS
REF-Level
-10 dBm
-10 dBm
-10 dBm
RF-Att. mode
AUTO
AUTO
AUTO
RF-Attenuator
20 dB
20 dB
20 dB
PRE-AMP
OFF
OFF
OFF
Sweep Points
501
4001
8001
Stop after sweep
OFF
OFF
OFF
Transd. factor
<NONE>
<NONE>
<NONE>
Limit check
ON
ON
ON
Limit
-13 dBm
-13 dBm
-13 dBm
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...