R&S FSP
Remote Control – Description of Commands
MMEMory Subsystem
6.140
Operating Manual 1164.4556.12 - 06
Parameter:
1 to 3 := selected measurement curve (trace)
<file_name> := file name in compliance to DOS conventions, including indication
of the path and the drive name
Example:
"MMEM:STOR2:TRAC 3,'A:\TEST.ASC'"
'Stores trace 3 from screen B in the file TEST.ASC on a disk.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
For Spurious Measurement an extended file format is used:
File contents
Explanation
File header
Type;R&S FSP;
Model
Version;3.9x;
Firmware version
Date;02.Feb 2007;
Storage date of data set
Mode;ANALYZER;SPURIOUS;
Operating mode of the device
Format for Spurious Emissions measurement
Start;9000.000000;Hz
Stop;8000000000.000000;Hz
Start/stop of the span
Unit: Hz
x-Axis;LIN;
Linear (LIN) or logarithmic (LOG) (future feature) scaling of the x-
axis
Sweep Count;1;
Selected number of sweeps
Range 1:
Loop across all defined sweep ranges (1 to 20)
Start; 9000.000000;Hz
Range start frequency in Hz
Stop; 150000.000000;Hz
Range stop frequency in Hz
Filter Type;NORMAL;
Range filter type: NORMAL, RRC or CFILTER
RBW;10000.000000;Hz
Resolution bandwidth of the measurement filter
VBW;30000.000000;Hz
Resolution bandwidth of the video filter
Auto Sweep time;ON
Sweep time; 0.145000;s
Selected sweep time in the current range
Detector;RMS;
Detector: MAX PEAK, MIN PEAK, RAVERAGE
REF-Level; -10.000000;dBm
Setting of the reference level in the current range
Auto RF-Attenuator; OFF;
Manual (OFF) or automatic (ON) setting of the RF attenuator
RF Att;15.000000;dB
Range input attenuation
Sweep Points;625;
Number of sweep points in the current range
Preamp; 0.000000;dB
Range preamplifier, on (20 dB) or off (0 dB)
Stop after range;OFF;
Stop after range, switched on (ON) or off (OFF) for the current
range
Transducer;TRD1;
Transducer name (if activated)
repeated for all sweep list ranges
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...