R&S FSP
Instrument Functions
Basic Settings
4.196
Operating Manual 1164.4556.12 - 06
SELECT ITEMS
The
SELECT ITEMS
softkey moves the selection bar to the first line, left column of
the
Items
field. An entry is selected. Position the entry focus to the corresponding
partial data set using the cursor keys and then press the
ENTER
key in the desired
line. The selection is cleared by pressing the key again.
The following submenu is opend:
Remote command:
MMEM:SEL:HWS ON
(Current Settings)
MMEM:SEL:LIN:ALL ON
(All Limit Lines)
MMEM:SEL:TRAC ON
(All Traces)
MMEM:SEL:SCD ON
(Source Cal Data)
MMEM:SEL:TRAN:ALL ON
(All Transducers)
Note that the command
MMEM:SEL:SCD ON
(Source Cal Data) is only available
with option R&S FSP-B09 or R&S FSP-B10.
The
Save
dialog box offers the following selectable data subsets in the
Items
field.
Current Settings
These settings include:
• current configuration of general instrument parameters
• current measurement hardware settings
• active limit lines:
A data set may contain maximum 8 limit lines for each
window. It always contain the activated limit lines and the
de-activated limit lines used last, if any. Consequently, the
combination of the restored deactivated limit lines depends
on the sequence of use with command
MMEM:LOAD
.
• the activated transducer factor
• user-defined color settings
• configuration for hardcopy output
• active transducer set:
A data set may contain maximum 4 transducer factors. It
always contain the activated factors and the factors used
and de-activated last, if any. Consequently, the combination
of the restored deactivated transducer factors depends on
the sequence of use with the command
MMEM:LOAD
.
• settings of tracking generator
(only with option R&S FSP B9)
All Limit Lines
all
limit
lines
All Transducer
all transducer factors
All Transducers
all
transducer
All Traces
all traces which are not blanked
Source Cal Data
correction data for tracking generator
(only with options R&S FSP B9 or R&S FSP B10)
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...