R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.98
Operating Manual 1164.4556.12 - 06
6.5.11
CALCulate:TLINe Subsystem
The CALCulate:TLINe subsystem controls the settings for the time lines. The measurement windows are
selected by CALCulate 1 (screen A) or 2 (screen B).
CALCulate<1|2>:TLINe<1|2>
0 to 1000s
This command defines the position of the time lines.
The time lines mark the times specified in the measurement window. Time lines are only available with
SPAN = 0.
Example:
"CALC:TLIN 10ms"
Characteristics:
*RST value: - (STATe auf OFF)
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:TLINe<1|2>:STATe
ON | OFF
This command switches the time line on or off.
Example:
"CALC:TLIN2:STAT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T