R&S FSP
Remote Control – Description of Commands
CALCulate Subsystem
6.48
Operating Manual 1164.4556.12 - 06
CALCulate<1|2>:MARKer<1...4>:MAXimum[:PEAK]
This command positions the specified marker to the current maximum value of the corresponding
trace. The corresponding marker is activated first or switched to the marker mode.
Example:
"CALC:MARK2:MAX"
'Positions marker 2 in screen A to the maximum value of 'the trace.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:RIGHt
This command positions the specified marker to the next smaller maximum value to the right of the
current value (i.e. in ascending X values) on the corresponding trace.
Example:
"CALC:MARK2:MAX:RIGH"
'Positions marker 2 in screen A to the next lower maximum value to the right of the
current value.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MINimum:AUTO
ON | OFF
This command switches an automatic minimum peak search for marker 1 at the end of each particular
sweep on and off. The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account. The numeric suffix at MARKer<1...4> is irrelevant.
Example:
"CALC:MARK:MIN:AUTO ON"
Activates the auto search function for marker 1.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
If no maximum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: -200) is produced.
If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...