Remote Control – Description of Commands
SENSe Subsystem
6.16 SENSe Subsystem
The SENSe subsystem is organized in several subsystems. The commands of these subsystems
directly control device-specific settings, they do not refer to the signal characteristics of the measure-
ment signal.
The SENSe subsystem controls the essential parameters of the R&S FSP. In accordance with the SCPI
standard, the key word "SENSe" is optional for this reason, which means that it is not necessary to
include the SENSe node in command sequences.
The measurement windows are selected by SENSe1 and SENSe2:
SENSe1 = Modification of screen A settings
SENSe2 = Modification of screen B settings.
Screen A is automatically selected if 1 or 2 is missing.
6.16.1
SENSe:AVERage Subsystem
The SENSe:AVERage subsystem calculates the average of the acquired data. A new test result is
obtained from several successive measurements.
There are two types of average calculation: logarithmic and linear. In case of logarithmic average calcu-
lation (denoted with VIDeo), the average value of the measured logarithmic power is calculated and in
case of linear average calculation, the linear power is averaged before the logarithm is applied.
The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
[SENSe<1|2>:]AVERage:COUNt
0 to 32767
This command defines the number of measurements which contribute to the average value.
It should be noted that continuous averaging will be performed after the indicated number has been
reached in continuous-sweep mode.
In single-sweep mode, the sweep is stopped as soon as the indicated number of measurements
(sweeps) is reached. Synchronization to the end of the indicated number of measurements is only
possible in single-sweep mode.
The command
[SENSe<1|2>:]AVERage:COUNt
is the same as command
[SENSe<1|2>:
]SWEep:COUNt
. In both cases, the number of measurements is defined whether the average
calculation is active or not.
The number of measurements is valid for all traces in the indicated measurement window.
Example:
"INIT:CONT OFF"
'Switching to single-sweep mode.
"AVER:COUN 16"
'Sets the number of measurements to 16.
"AVER:STAT ON"
'Switches on the calculation of average.
"INIT;*WAI"
'Starts the measurement and waits for the end of the 16 sweeps.
Characteristics: *RST value:
0
SCPI: conform
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...