Operating Manual 1164.4556.12 - 06
4.111
R&S FSP
Instrument Functions
Analyzer Mode
• y-axis lower limit: 1E-6
Remote command:
CALC:STAT:PRES
MEAN PWR
POSITION
Defines the relative X position of the mean power value for the CCDF measurement.
The default position is 0% (left corner of the grid). This softkey is available if the
CCDF measurement is switched on.
Remote command:
CALC:STAT:SCAL:X:MPOS <numeric_value>
ADJUST
SETTINGS
“ADJUST SETTINGS” on page 4.110
CONT MEAS
The
CONT MEAS
softkey starts collecting a new sequence of sample data and cal-
culating the APD or CCDF curve depending on the selected measurement. The next
measurement is started automatically as soon as the indicated number of samples
has been reached.
Remote command:
INIT:CONT ON;
INIT:IMM
SINGLE MEAS
The
SINGLE MEAS
softkey starts collecting a new sequence of sample data and
calculating the APD or CCDF curve depending on the selected measurement. At the
beginning of the measurement previously obtained measurement results are dis-
carded.
Remote command:
INIT:CONT OFF;
INIT:IMM
Hint for usage of the marker functions with measurement of signal
statistics
With the signal statistic measurement level always is displayed on x-axis. Y-axis
always is a normalized value between 0 and 1. In contrary to use of marker in fre-
quency or time domain marker is input in level values and the output is in percent-
age values.
Example
Measurement of CCDF of a IS95 BTS signal, level 0 dBm, frequency 800 MHz
[
PRESET
]
Switch on preset settings.
[
FREQ
:
CENTER
: 800
MHz]
Set center frequency to 800 MHz.
[
AMPT
: 10 dBm]
Set reference level to 10 dBm.
[
BW
: 3 MHz]
Set resolution bandwidth to 3 MHz (resolution bandwidth
shall be wider then signal bandwidth (1.25 MHz) in order
to have the complete signal within the resolution band-
width).
[
MEAS
]
Call the menu for measurement functions.
[
SIGNAL STATISTIC
]
Call the menu for signal statistics measurement.
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...